Schedule 1262 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Description:
Radium Isotope |
|
Analyzing
Laboratory(s): |
|
Eberline
Services, Richmond, CA . |
|
|
Analyte |
|
CAS Number |
RL |
Unit |
|
***
Radium-226 |
|
13982-63-3 |
0.1 |
pCi/L |
|
***
Radium-228 |
|
15262-20-1 |
1 |
pCi/L |
|
|
Container
Requirements |
|
Quantity |
Bottle |
|
1 |
2L FAR |
|
|
Description: 1 L Polyethylene
bottle, acid-rinsed |
|
|
Treatment and Preservation: Filter
through 0.45-um filter and acidify with Ultrex nitric acid (HNO3) to pH <
2 |
|
References |
|
|
1. EPA 903.1 |
|
Radium-226
in drinking water - Radon emanation technique |
|
Method ID:
EPA 903.1 |
|
|
2. EPA 904.0 - Modified |
|
Radium-228
in drinking water (with modification) |
|
Method ID:
EPA 904.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Lab Code 1369 |
|
Description:
Radon-222, Water, Unfiltered, Time Sensitive!! MUST be rcvd within 48 hrs of
collection, Liq Scint |
|
Analyzing
Laboratory: USGS - National Water Quality Laboratory, Denver, CO |
|
|
Parameter Name |
|
CAS Number |
RL |
Unit |
|
Radon-222 |
|
14859-67-7 |
26 |
pCi/L |
|
Radon-222,
2 sigma precision estimate |
|
pCi/L |
|
Radon-222,
minimum detectable concentration |
|
pCi/L |
|
|
Container
Requirements |
|
20mL RURCV |
|
Description: |
|
Treatment
and Preservation: 20 mL glass vial filled with 10 mL mineral oil liquid
scintillation cocktail and 10 mL of sample. Time Dependent - ship immediately
- MUST be received within 48 hours of collection. The vials are ordered from
One Stop shopping, stock number 1390. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Lab Code 1565 |
|
Description:
Tritium, Water, Unfiltered, by Electrolytic Enrichment and Liquid
Scintillation |
|
Analyzing
Laboratory: USGS-NATIONAL RESEARCH PROGRAM LAB, MENLO PARK, CA |
|
|
Parameter Name |
|
CAS Number |
RL |
Unit |
|
|
Tritium |
|
10028-17-8 |
1 |
pCi/L |
|
tritium, 2
sigma precision estimate |
|
pCi/L |
|
|
Container
Requirements |
|
1L RUR |
|
Description:
250 or 500 or 1000 mL Polyethylene bottle, Must use polyseal cap. Use
unfiltered sample to rinse bottles. For labcodes 452 |
|
Treatment
and Preservation: 460, 624, and 1565: see NWQL tech memo 95.11 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Lab Code 1043 |
|
Description:
Tritium, Water, Unfiltered, by Electrolytic Enrichment and Gas Counting |
|
Analyzing
Laboratory: University of Miami, Tritium Laboratory, Miami, FL |
|
|
Parameter Name |
|
CAS Number |
RL |
Unit |
|
|
Tritium |
|
10028-17-8 |
0.3 |
pCi/L |
|
tritium, 2
sigma precision estimate |
|
pCi/L |
|
|
Container
Requirements |
|
1L RUR |
|
Description:
250 or 500 or 1000 mL Polyethylene bottle, Must use polyseal cap. Use
unfiltered sample to rinse bottles. For labcodes 452 |
|